Patient Monitor PM-2000M

PT Sano Medika Indonesia

Patient Monitor PM-2000M

Technical Specifications

Safety Standards Safety Standars IEC 60601-1+A1+A2; IEC 60601-2+A1; IEC 60601-1-8 (Alarm)
Physical Specifications Dimensions Weight 17” 425mm (L) x 245mm (W) x 384mm (H) / 15” 384mm (L) x 213mm (W) x 320mm (H)

17 ” Standard configuration 14kg / 15” 7,5 kg Includes TM module

Display Display

Resolution Waveforms Displayed Selectable Interface

17” or 15” Color TFT LCD Touch-screen 1280X1024dpi

Up to 15

Standard screen monitoring, Trend / Monitoring , Co – Display, bed to bed view (optional ) OxyCRG screen with dynamic view, calculation of drug dosages interface.

Environment Requirement Ambient Temperature Humidity -20 ºC ~ 55ºC (-4 ~ 131ºF)

15% – 95% without condensation

Energy Supply Power Supply Internal 100-240V CA, 50/60HZ
  Battery Battery Rechargeable Li -Ion 4200 mAh
  Working Period 2 hours max. (2 batteries 4200mAh)
  Recharging Time ≤350 minutes (4200 mAh)
Resp Method Impedance between RA- LL , RA -LA
  Operation Mode Auto / Manual
  Rr Measurement Range Adult: 0-120rpm / Neonate-Pediatric: 0-150rpm
  Resolution 1rPM
  Apnea Alarm Threshold 10s, 15s, 20s (default), 25s, 30s, 35s, 40s
  Alarm 3 levels of audible and visual alarm, Alarms events recallable
  Band Width 0.2-2.5Hz (-3dB)
  Sweep Speed 6.25mm/s, 12.5mm/s, 25mm/s, 50mm/s
ECG Lead Type

Gain Selection Sweep Speed ECG Hr Range

Resolution & Accuracy Filter

St-Segment Detection

3 – Leads: I, II , III / 5 – Leads: I, II , III, aVR , aVL, aVF, V / 12, Derivations : I, II , III , aVR, aVL, aVF, V1, V2, V3 , V4, V5 , V6

x0.125, x0.25, x0.5, x1, x2, x4

6.25mm/s, 12.5mm/s, 25mm/s, 50mm/s

Adult:15-300bpm / / Pediatric-Neonate: 15-350bpm

±1bpm or ±1%, whichever is greater

Diagnosis mode: 0.05-150Hz / Monitoring mode: 0.5-40Hz / Surgical mode: 1-20Hz Measurement range: -2.0mV ~ 2.0mV / Alarm range: -2.0mV ~ 2.0mV

Comes with St-segment Arrhythmia Analysis and Pacemaker detection

NIBP Method  Operations Modes Auto Measurement Time Interval Measurement Unit

Measurements Types Pressure Range Adults Pressure Range Pediatrics Pressure Range Neonates Resolution

Accuracy Alarm

Pr From PANI Resolution Accuracy

Oscillometric

Manual / Automatic/ Continuous 1/2/3/4/5/10/15/30/60/90/120/240/480 minutes.

 

MmHg/kpa Selectable Systolic, Diastolic, Mean

Systolic: 40 – 270 mmHg / Diastolic: 10 – 215 mmHg / Mean: 20 – 235 mmHg

Systolic: 40 – 200 mmHg / Diastolic: 10 – 150 mmHg / Mean: 20 – 165 mmHg

Systolic: 40 – 135 mmHg / Diastolic: 10 – 100 mmHg / Mean: 20 – 110 mmHg 1mmHg

Max mean error ±5mmHg / Max standard deviation ±8mmHg Systolic, Diastolic, Mean

Measurement 40-240 bpm

1 bpm

±3bpm or 3.5% whichever is greater

Comes with leak test and pressure auto calibration

NIBP (Optional OMRON® M3600) Measurement Range Adults/Pediatric Neonates

Accuracy

OMRON® is registered mark

Systolic: 60 – 250 mmHg / Diastolic: 40 – 200 mmHg / Mean: 45 – 235 mmHg

Systolic: 40 – 200 mmHg / Diastolic: 20 – 90 mmHg / Mean: 30 – 100 mmHg

±2bpm or 2% of the reading

SPO2 Measurement/Alarm Range 0 – 100% (SpO2)
  Resolution 1%
  Accuracy ±2% (70-100% Adult/Pediatric); / ±3% (70-100% Neonate)
  PR Measurement 25 ~ 300 bpm
  Alarm Range 30 ~ 300 bpm
  Resolution 1bpm
  Accuracy ±2bpmm
  Refresh 1s
SPO2 (Optional,      Measurement/Alarm Range 20 ~ 300bpm

Nellcor Oximax™)  Resolution                             1bpm

Accuracy                                 ± 3bpm (20bpm ~ 250bpm)

Temperature                                 Measurement (2 Channels, 1       Resolution

Probe By Default) Range Accuracy

0-50ºC (32-122ºF)

0.1ºC

±0.1ºC (without probe)

IBP                         Technology Measurement Resolution Accuracy Sensitivity Impedance Range Filter Invasive direct measurement

ART, PA, CVP, RAP, LAP,ICP, P1, P2

1 mmHg

± 2% or ± 1 mmHg (whichever is greater) 5µ V/V / mmHg

300 ~ 3000 Ω

DC ~ 12.5Hz; DC ~ 40Hz

BIS

Primary Parameter Secondary Parameter

Scan Speed Wave Scale

Bis Trend Length Attenuation Range EEG Band Width BIS Alarm Range

Bispectral index, power spectrum analysis, measurement parameters BIS 0 ~100

SQI/SR 0% ~100% / EMG 30dB ~80Db / SEF 0.5Hz ~ 30.0Hz / TP 40dB ~100dB / BC

(applicable only to BIS™ Extend Sensor ) 0 ~ 30 6.25mm/s,12.5mm/s, 25mm/s, 50mm/s

50µv, 100µv, 200µv, 500µv 6min, 12min, 30min, 60min 10s, 15s, 30s

0.25Hz~50Hz

0~100

CO2                                    Technology Rango EtCO2

Precision

Infrared Absorption 0~ 150mmHg

± 2% – 0~ 40mmHg / ±5% – 41~70mmHg / ±8% – 71~100mmHg / ±10% – 101~150 mmHg /

± 12% of reading, FR more than 80 rpm (sidestream)

C.O.                        Method

Measurement Range TB Alarm Range

Thermodilution technology

CO: 0.1~ 20L/min / TB: 23ºC~43ºC (73.4°F~109.4°F) / TI: -1ºC~27ºC (30.2°F~80.6°F) 23ºC~43ºC (73.4°F~109.4°F)

Anesthetic Gas/O2 Technology

Paramagnetic Oxygen Gas

Warm-Up Time

Infrared Absorption Optional

CO2, O2, N2O, Des, Iso, Enf, Hal, Sev

(IRMA AX+) Precision mode ISA : 10s / mode full precision: 20s / ( ISA OR + / AX + ) < 20s / Sampling flow ( for ISA or + / AX + ) 50 ± 10 ml / min

MAC Visualization of the value

Printer

 

Printing Speed Paper Width

Dot Matrix printer

You can record up to three waveforms, selectable by the user. Waveform recording and real-time

12.5mm/s, 25mm/s, 50mm/s 50 mm

I/O Interface         Peripherals 4-USB Port
  SD card socket
RS-232 serial port
RJ-45 ethernet port, IEEE 802.3
VGA output
Analog and nurse call output
Defibrillation synchronization output
PAM connector
DVI connector
Network interface